HI-6222 pH/ORP/ISE Multiparameter Modular Benchtop System
The HI-6222 Kit includes the HI-6000 modular multiparameter meter system and two of the HI-6000-2 pH/ISE modules. Complete with the HI-1131B pH electrode and HI-7662-TW temperature probe and calibration solution starter set for pH providing everything you need for accurate measurements of pH, ORP and ISE.
HI-6000 is a fully flexible measurement platform that can be customised according to the user’s application needs. The kit is compatible with a wide range of electrodes for ISE measurement and can be combined with other modules simultaneously, allowing for measurement of a wide range of samples at once.
Modules available for the HI-6000 include:
- HI-6000-1 module for pH/ORP
- HI-6000-2 module for pH/ORP/ISE
- HI-6000-3 module for EC
- HI-6000-4 module for DO
Contact our team for advise on choosing the right module for your needs and expand the functionality of your kit.
Accurate and Reliable
Together with appropriate electrodes, the meter provides highly accurate and reliable measurement for all laboratories. Up to three measurement channels can be viewed simultaneously. Views can be mixed and matched. Data can be shown in basic or in full detail, in a table or plotted as a graph.
The HI-6000 Series allows for up to 5-point calibration for pH, ISE and Conductivity for optimum results.
- pH calibration up to 5 points with 8 stored standard buffers and 5 custom buffers
- ISE calibration up to 5 points with 5 stored standards and 5 custom buffers
- Conductivity calibration up to 5 points with 6 stored standards and 1 user standard
Storage Capacity & Data Sharing
Stores up to 1,000,000 data points for each user.
- Data can be shared via Wi-Fi, Ethernet, email, web server, USB flash drive or PC
- No software is required
- Conveniently connect a keyboard, a barcode scanner, a printer or two stirrers to improve your workflow
Storage Capacity & Data Sharing
| Specifications of HI-6222-02 pH/ORP/ISE Kit | |
|---|---|
| Range | -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH |
| Resolution | 0.1 pH; 0.01 pH; 0.001 pH |
| Accuracy | ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
| Temperature Compensation | Automatic or manual |
| _____ | |
| mV | |
| _____ | |
| Range | -2000.0 mV to 2000.0 mV |
| Resolution | 1 mV; 0.1 mV |
| Accuracy | ±0.2 mV ±1 last significant digit |
| _____ | |
| ISE | |
| _____ | |
| Range | 1 x 10-6 to 9.99 x 1010 concentration |
| Resolution | 1; 0.1; 0.01; 0.001 concentration |
| Accuracy | ±0.5% (monovalent ions); ±1% (divalent ions) |
| Calibration | Automatic, up to five-point calibration, seven fixed standard solutions available for each measurement unit, and five user defined standards |
| _____ | |
| Temperature | |
| ______ | |
| Range | -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.0 to 393.0 K |
| Resolution | 0.1 °C / 0.1 °F / 0.1 K |
| Accuracy | ±0.2 °C / ±0.4 °F / ±0.2 K |
| Relative mV offset range | ±2000.0 mV |
| _____ | |
| pH Calibration | |
| _____ | |
| Calibration Points | Up to 5 |
| Type | Automatic; Semiautomatic; Manual |
| Standard Buffers | pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 |
| Custom buffers | up to 5 |
| Custom Group | up to 5 |
| 1st calibration point | Offset or Points (user setting) |
| Temperature user calibration | Single point, adjustable; Isopotential point: -2.000 to 20.000pH |